11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-70
Continued
No.
Message
966
Cleaning –
Additive
will expire
soon
967
Creatine
high.
Unable to
perform
Crea
measure-
ment
969
FLEXQ
module not
expected
970
Rilibäk
violation:
Value above
upper limit
971
Rilibäk
violation:
Value belov
lower limit
ABL800 FLEX Operator's Manual
Interpretation
The Cleaning additive
will expire soon.
The Creatine
concentration is too
high compared to the
Crea concentration and
the analyzer is not able
to measure Crea within
the specifications.
See chapter 14:
Specifications in the
Operator's Manual
Software configuration
does not match
analyzer type.
Operator action
Replace the solution before
the lifetime is actually
exceeded.
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform a High Crea
Check. If the High Crea
Check result is accepted,
the blood sample may be
suspected.
- Perform measurement on
a new blood sample.
- Perform the Electrode
Troubleshooting procedure
Contact Radiometer
Service Representative.
Removal condition:
Matching software and
hardware configuration.
Continued on next page