11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-36
Continued
No.
Message
680
pH/BG
module not
active
El/Met
681
module not
active
682
Oxi module
not active
683
Inlet
module not
active
ABL800 FLEX Operator's Manual
Interpretation
The pH/BG module is
not responding due to
internal communication
problem.
The El/Met module is
not responding due to
internal communication
problem, or software
configuration does not
match the analyzer
type.
The Oxi module is not
responding due to
internal communication
problem, or software
configuration does not
match the analyzer
type.
The inlet module is not
responding due to
internal communication
problem.
Operator action
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition:
pH/BG module ready.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition: El/Met
module ready, or software
configured without El/Met
module support.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition: OXI
module ready, or software
configured without OXI
module support.
- Shut down power using
the Temporary Shutdown
function, then restart the
analyzer.
- Contact Radiometer
Service Representative.
Removal condition: Inlet
module ready.
Continued on next page