RADIOMETER ABL800 FLEX Operator's Manual page 138

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

3. Installation and setup
Communications,
LIS/HIS
Step
Connection
Setup
(continued)
This program allows you to connect the analyzer to the QA portal.
QA Portal
If the QC Portal communication is enabled, the analyzer will automatically send
QC results and Cal Verification measurements to the QA portal.
To enable communication with the QA portal, do the following:
Step
3-72
Continued
6.
a. Connection specifications for serial low-level protocol:
Use the up and down arrows in each box to select baud rate, stop bits,
data bits and parity. Port number should be COM2.
Box
Option
Baud rate
1200, 2400, 4800, 9600, 14400, 19200, 38400, 115200
Data bits
5, 6, 7, 8
Stop bits
1, 1.5, 2
Parity
Even, odd, none.
b. Connection specifications for network low-level protocol:
Touch the screen to highlight the following boxes one after another:
Server address − Port number − Reconnect Interval (time interval in
seconds to attempt re-connection if connection failed) – Retransmit
max – Transmit timeout.
Use the keypad/keyboard to enter the relevant information.
c. Connection specifications for POCT1-A DML low-level protocol:
Touch the screen to highlight the following boxes one after another:
Server address − Port number − Reconnect Interval.
Use the keypad/keyboard to enter the relevant information.
Action
1.
Touch and highlight the "Server address" box.
Type in the TCP/IP address of your QA portal, using the screen
keypad or keyboard.
ABL800 FLEX Operator's Manual