ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
862
pH-only
sample
handling
adapted to
35uL range
864
pH-only
sample
handling
adapted to
55uL range
865
pH/BG
sample
handling
adapted to
35uL range
(decay to
pH only)
867
pH/BG
sample
handling
adapted to
55uL range
868
pH/BG/Oxi
sample
handling
adapted to
55uL range
Interpretation
To counteract
insufficient or
inhomogeneous
sample, the analyzer
has automatically
adjusted the sample
handling to the sample
volume.
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
To counteract
insufficient or
inhomogeneous
sample, the analyzer
has automatically
adjusted the sample
handling to the sample
volume.
11. Troubleshooting
Operator action
None.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
None.
Continued on next page
11-53