11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-18
Continued
No.
Message
379
Calibration
unstable.
(response
fault)
386
Barometer
out of range
408
Zero current
error
443
Ca(7.4) not
usable
ABL800 FLEX Operator's Manual
Interpretation
An electrode response
fault occurred during
calibration.
Affected parameters
and measurements
(only creatinine) are
marked with "?".
The barometer reading
is out of the range: 450-
800 mmHg or 60.0-
106.7 kPa.
Measurement results
are marked with "?".
Zero current was too
high during calibration.
Affected parameters
and measurements are
marked with "?".
cCa++ at a pH of 7.4 is
not usable as the actual
pH is out of the range
7.2-7.6.
Operator action
- Check for and remedy
any System Messages.
- Perform any pending
replacements including
electrodes.
- Check that electrodes are
properly installed.
- Verify that proper
solutions and gases are
used.
- Perform the Electrode
Troubleshooting procedure.
Removal condition:
Successful 1- or 2-point
calibration.
- Check the measured
pressure in Barometer
Setup program and
compare it to that of an
external barometer. Adjust
the analyzer pressure
accordingly.
- Contact Radiometer
service representative.
- Check for and remedy
System Messages.
- Perform any pending
replacements, including
electrodes.
- Perform the Electrode
troubleshooting procedure.
- Contact Radiometer
service representative.
No action required.