13. Specifications
Measured parameters,
Pleural liquid
* If the measured values obtained lie outside the test range, Radiometer advises
you to repeat the measurement by means of another method.
Expired air
Baro
Other fluids
All parameters available on the analyzer.
13-4
Continued
Parameter
Unit
pH
pH scale
Parameter
Unit
pCO
mmHg
2
kPa
pO
mmHg
2
kPa
Parameter
Unit
p(amb)
mmHg; Torr
kPa
ABL800 FLEX Operator's Manual
Measuring Range
6.300-8.000
Measuring Range
5.0-250
0.67-33.3
0.0-800
0.00-107
Measuring Range
450-800
60.0 -106.7
Test range
7.0-7.5*
Test range
15-150
2.00-20.00
15-530
2-70
Test range
450-800
60.0-106.7