RADIOMETER ABL800 FLEX Operator's Manual page 343

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
920-922
Continued
No.
Message
Module not
filled within
time limit in
wash with
cleaning
923
Inlet not
refilled
properly
during bi-
directional
pump
calibration
924
Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 1
925
Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 2
Interpretation
Module did not fill
properly with cleaning
solution.
The inlet module was
not filled properly
during bi-directional
pump calibration.
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
11. Troubleshooting
Operator action
- Check the solution
containers for adequate
volume. Replace as
required.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful Rinse
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure
Removal
condition:Successful pump
calibration.
Continued on next page