Fluid transport system description,
Rinse process
Stage
(continued)
Carefully observe the flow of fluid through the system during the rinse. Use the
Evaluating the
table below to help identify possible sources of error.
rinse
Possible cause
Obstruction in
inlet
Improper
electrode
installation
Obstruction in
the measuring
chamber of a
measuring
module
3
The measuring modules are filled with rinse solution flowing through
the measuring chambers from the bottom up.
The pH/BG module is filled with gas containing CO
4
from the top, expelling the rinse solution from the measuring
chambers.
5
The outer portion of the inlet system is filled with air drawn from the
exterior (not shown).
• Fluid flowing upwards through
pH/BG, El/Met and Met II
modules during Stage 1.
• Erratic flow of fluid through all
three modules that contains
excessive air bubbles.
• Fluid does not move through the
small canals off the waste pump
tubes, just above the pump rotor.
• Excessive suction noise coming
from inlet.
• Fluid and air bubbles rising up
into a measuring chamber.
• Leakage from electrode
compartment.
Fluid does not flow freely through all
measuring chambers of a particular
measuring module.
Continued
Description
Indication
11. Troubleshooting
. The gas is filled
2
Recommended
action
Perform inlet
troubleshooting
procedure.
Ensure the electrodes
are properly
installed.
Clean the measuring
chambers of the
measuring module.
11-83