11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-46
Continued
No.
Message
766
ABL not
connected to
RADIANCE
– no
RADIANCE
connection
license
767
ABL not
connected to
RADIANCE
– ABL
StatLink
version too
high
768
ABL not
connected to
RADIANCE
– ABL
StatLink
version too
low
770
Failed to
restore
Custom
Setup
771
Succeeded to
restore
Custom
Setup
772
User
Activity:
773
Remote
operator
logged on
with user:
ABL800 FLEX Operator's Manual
Interpretation
The analyzer has been
refused connection to
RADIANCE because
there is no connection
licence available on
RADIANCE.
The analyzer has been
refused connection to
RADIANCE because
the ABL StatLink
version is higher than
the RADIANCE
StatLink version.
The analyzer has been
refused connection to
RADIANCE because
the ABL StatLink
version is lower than
the RADIANCE
StatLink version.
The setup could not be
restored.
Restoring of setup is
completed.
User activity logged by
operator.
A remote operator has
logged on the analyzer
via NetOp
Operator action
Contact RADIANCE /IT
engineer or Radiometer
service representative.
Removal condition:
Connection to RADIANCE
established.
Contact RADIANCE /IT
engineer or Radiometer
service representative.
Removal condition:
RADIANCE connection
established.
Contact RADIANCE /IT
engineer or Radiometer
service representative.
Removal condition:
RADIANCE connection
established.
- Download the setup data
from another floppy disk,
hard disk or network.
- Contact Radiometer
service representative if the
error persists.
No action required.
No action required.
No action required.
Continued on next page