RADIOMETER ABL800 FLEX Operator's Manual page 291

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
128
LIS/HIS:
Failed to
open
connection
129
LIS/HIS:
Failed to
close
connection
131
LIS/HIS:
Failed to
send packet
132
LIS/HIS:
Failed to
receive
packet
133
LIS/HIS:
Connection
lost
Interpretation
The communication
hardware was busy or
the remote system did
not respond.
Messages were queued
when the communica-
tion channel was
closed. Results and
other messages sent by
the analyzer to a
remote system may be
lost.
A communication error
occured while sending
a message. The
message was not sent.
An error occured while
receiving a message.
The analyzer was not
able to recognize the
received massage.
A previously
established LIS/HIS
connection has been
lost.
11. Troubleshooting
Operator action
- Check that the remote
system is running, correctly
configured and responding.
- Check communication
parameters, e.g. baud rate,
parity, IP address, etc., as
defined in Communication
setup.
- Reboot the analyzer.
If the problem persists,
check the communication
hardware. The remote
system may lack buffer
capacity.
- Check that the remote
system is running and
responding.
- Check communication
hardware including cables.
- Repeat sending.
- Check that protocol types
are correctly configured on
both the analyzer and the
remote system.
- Contact Radiometer
service representative.
- Check that the remote
system is running and
responding.
- Check cables.
Continued on next page
11-9

Advertisement

Table of Contents
loading