7. Replacements
Cleaning the analyzer,
When cleaning the analyzer covers and outer case, use soapy water or a mild
Cleaning the
detergent.
analyzer
exterior
Do not use abrasive cleansers or pads, ethanol-based substances or aggressive
detergents for cleaning.
Cleaning the
Use a dry or lightly dampened soft lint-free cloth to clean the analyzer's touch
analyzer touch
screen. Wipe the screen gently to remove fingerprints and/or dirt. To avoid
screen
streaking, an approved screen cleaner is recommended.
Cleaning the
Use a tissue lightly dampened in mild soapy water.
FLEXQ sampler
If you wish to clean the sampler tray cover separately, do the following:
tray cover
Step
Disinfection of
Disinfection of outer surfaces is performed when appropriate. Disinfection
outer surfaces
frequency depends on local requirements and the use of the analyzer.
Prior to disinfection always ensure that that analyzer surfaces are clean and without
residues from blood and/or liquids.
Use the following disinfectants:
Wipe the outer surfaces of the analyzer and the touch screen, using disinfectant on
a paper towel or tissue.
7-24
Continued
Action
1.
2.
Use a tissue lightly dampened in mild soapy water to clean it.
3.
Remount the sampler tray cover
CAUTION – Safety precautions
Follow legal and local rules for safe work practices with chemicals.
70 % isopropyl alcohol
70 % ethanol
4 % Diversol BX