RADIOMETER ABL800 FLEX Operator's Manual page 340

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-58
Continued
No.
Message
890
Error
measuring
Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
892
Inlet not
filled
properly
during
pump
calibration
using
reagent as
backup
893
Sample in
Met II's
upper liquid
sensor
detected
prematurely
during
aspiration
ABL800 FLEX Operator's Manual
Interpretation
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
During pump
calibration
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
Met II module was not
filled correctly during
sample aspiration.
Operator action
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
- Ensure sample is free of
clots.
- Repeat measurement.
- Perform the Fluid
Transport Troubleshooting
procedure.
Continued on next page