11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-64
Continued
No.
Message
937
Cleaning
Solution -
Crea will
expire soon
938
Cal1
Solution -
Crea is
expired
939
Cal2
Solution -
Crea is
expired
940
Cleaning
Solution -
Crea is
expired
ABL800 FLEX Operator's Manual
Interpretation
The additive used in
the Cleaning Solution
to measure the High
Crea Check will expire
soon.
This is a warning - the
solution lifetime is 14
days.
The additive used in
the Calibration
Solution 1 to calibrate
Crea is expired.
The solution is more
than 14 days old.
It is no longer possible
to calibrate Crea.
The additive used in
the Calibration
Solution 2 to calibrate
Crea is expired.
The solution is more
than 14 days old.
It is no longer possible
to calibrate Crea.
The additive used in
the Cleaning Solution
to measure the High
Crea Check is expired.
The solution is more
than 14 days old.
It is no longer possible
to perform the High
Crea Check.
Operator action
Replace the solution before
the lifetime is actually
exceeded.
Replace the Calibration
Solution 1 bottle -
remember to add the
additive.
Removal condition:
Fresh bottle installed.
Replace the Calibration
Solution 2 bottle -
remember to add the
additive.
Removal condition:
Fresh bottle installed.
Replace the Cleaning
Solution bottle - remember
to add the additive.
Removal condition:
Fresh bottle installed.
Continued on next page