RADIOMETER ABL800 FLEX Operator's Manual page 297

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
339-340.
342-343,
345-346
Continued
No.
Message
328
No leading
air segment
in inlet's
liquid
sensor
within time
frame
329
QC
expiration
date
exceeded
331
No sample
detected
during
sample
aspiration
332,
pH/BG
module not
333
filled
Pump
calibration
error
Interpretation
Air segment in front of
sample is too small or
missing during sample
aspiration, possibly due
to obstructions in inlet
or problems with liquid
sensors. Measurement
is aborted.
The quality control
measurement was
performed on an
expired control
solution.
No sample detected in
the pH/BG module.
Measurement is
aborted.
Sample did not fill the
pH/BG module
properly during
aspiration.
Measurement is
aborted.
The pump calibration
failed for one or more
pumps. The previous
pump calibration is
used. A successful
pump calibration
should be attempted as
soon as possible.
11. Troubleshooting
Operator action
- Check inlet for
obstructions and clean as
required.
- Perform liquid sensor
adjustment.
- Perform the Fluid
Transport Troubleshooting
procedure.
Discontinue the use of the
lot and set up a valid lot for
the control solution.
- Ensure that adequate
sample volume is used.
- Check the sample for
clots.
- Perform the Fluid
Transport Troubleshooting
procedure.
- Ensure that adequate
sample volume is used.
- Check the sample for
clots.
- Perform the Fluid
Transport Troubleshooting
procedure.
- Check rinse solution for
sufficient volume. Replace
as necessary.
- Repeat pump calibration.
- Perform the Pump
Troubleshooting procedure.
Removal condition:
Successful pump
calibration.
Continued on next page