ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
688
ctHb/ceHb
too low for
OXI
calculation
Gas
689
pressure
offset
voltage out
of range.
692
ABL not
connected to
RADIANCE
693
ABL not
connected to
RADIANCE
- incorrect
password.
694
ABL
connected to
RADIANCE
Interpretation
ctHb < 1 mmol/L or
ceHb < 0.75 mmol/L. If
ctHb is too low FHHb,
FO
Hb, FCOHb and
2
FmetHb are not
calculated. If ceHb =
cHHb + cO
Hb is too
2
low sO
is not
2
calculated.
The gas pressure offset
voltage measured by
the transducer is out of
range: 0.07-0.35 V.
The analyzer is not
connected to
RADIANCE.
The analyzer was
refused connection to
RADIANCE due to
incorrect password.
The analyzer is
connected to
RADIANCE.
11. Troubleshooting
Operator action
Informative message. No
analyzer error detected.
- Restart analyzer.
- Contact Radiometer
service representative.
Removal condition: Gas
pressure transducer offset
within the required range.
Contact your RADIANCE
/IT engineer.
- Check RADIANCE
communication setup
including TCP/IP address,
port No. and password.
- Check that RADIANCE
is responding.
- Check network
connections.
Removal condition:
RADIANCE connection
established or disabled.
Enter the correct password
in the analyzer's
RADIANCE
Communication setup.
Removal condition:
RADIANCE connection
established or disabled.
No action required.
Continued on next page
11-37