RADIOMETER ABL800 FLEX Operator's Manual page 412

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

15. Radiometer settings
Radiometer default settings,
Analysis Setup
defaults
(continued)
Default settings for Drift tolerances are as follows:
Calibration
Setup defaults
15-6
Temperature Corrected Values
pH(T)
pCO
(T)
2
pO
(T)
2
< New Group >
Oxygen Status
ctO
2
p50
< New Line >
Acid Base Status
cBase(Ecf)
(
P,
c
HCO
3
Parameter
Drift 1
pH
pCO
2
pO
2
ctHb
+
cK
+
cNa
2+
cCa
-
cCl
cGlu
cLac
cCrea
ABL800 FLEX Operator's Manual
Continued
st
)
Drift 2
0.020
0.020
2.5
5
6
6
10
N/A
0.2
1.5
3
1
0.05
0.2
2
3
0.5
N/A
0.2
N/A
15
15
Units
mmHg
mmHg
pA
mmol/L
mmol/L
mmol/L
mmol/L
mmol/L
mmol/L
mol/L
Continued on next page