Handler Interface
Using the Handler Interface
Setting Up the Handler Interface Board for Use with the List Sweep
Comparator
Follow these steps to set up the handler interface for the list sweep
comparator:
Step 1. [Meas Setup] - LIST SETUP to open the LIST SWEEP SETUP page.
Step 2. Set up the list table (by defining the sweep points and upper/lower limits) so
that you can conduct list sweep measurement. For more information, see the
chapter on "Configuring Measurement Conditions."
Step 3. Press the [System] key to open the SYSTEM INFO page.
Step 4. Using the cursor movement keys on the front panel, move the cursor to the
HANDLER I/F field. The following softkeys appear on the softkey label area.
— ON
— OFF
Step 5. Press the ON softkey. The handler interface is enabled for signal
input/output.
Step 6. Press the [Display Format] key to open the MEAS DISPLAY page. Press the
LIST SWEEP softkey so that you can begin list sweep measurement.
Keysight E4980A/AL User's Guide
-481