Measurement Procedure and Examples
Capacitor Measurements
2. Press the [Meas Setup] key and the CORRECTION softkey.
3. Move to the OPEN field by using the cursor keys and press the MEAS OPEN
4. Press the ON softkey to enable an OPEN correction.
5. Connect a shorting bar to the Keysight 16047E to set up the SHORT
Figure 7-19
Connecting a shorting bar
6. Move to the SHORT field using the cursor keys and press the MEAS SHORT
7. Press the ON softkey to enable a SHORT correction.
Step 5. Connect DUT to the test fixture as shown in
212
softkey.
Wait until the message "OPEN measurement in progress" disappears.
condition as shown in
softkey.
Wait until the message "SHORT measurement in progress" disappears.
Figure
7-19.
Figure 7-20
Keysight E4980A/AL Precision LCR Meter