Overview of Remote Control
Trigger System
When starting the trigger system with one of the following commands, the
system switches from "Idle" to the "Waiting for Trigger" state. The conditions of
transition differ depending on which trigger source is specified (Refer to
8-14 on page 249
When the trigger source is set to the external trigger (EXT), the conditions for
switching from "Idle" to "Waiting for Trigger" differ between the remote mode
and local mode.
Remote mode
Local mode
When the trigger source is set to the internal trigger (INT) or manual trigger
(MAN), the trigger system automatically switches from the "Idle" state to the
"Waiting for Trigger" state.
The E4980A/AL in the "Idle" state switches to the "Measurement" state by
using the :TRIG command.
When the :ABOR command is executed, the trigger system switches to the
"Idle" state regardless of which state the E4980A/AL is currently in. An
error occurs when query commands of the :FETCh subsystem are executed
because no measurement data exist.
"Waiting for Trigger" State
When the trigger system is in the "Waiting for Trigger" state and either the
instrument is triggered (i.e., a trigger is detected) or one of the following
commands is executed, the trigger system switches to the "Measurement"
state.
As shown in the table below, the instrument is triggered differently depending
on which trigger source is specified. To specify the trigger source, use the
:TRIG:SOUR command.
Trigger Source
Internal trigger
(Internal)
246
and
Figure 8-15 on page
:INIT[:IMM]
:INIT:CONT
When the :INIT:CONT is ON, the trigger system automatically
switches to "Waiting for Trigger."
When the :INIT:CONT is OFF, the trigger system switches to
the "Waiting for Trigger" state by sending the :INIT:IMM
command.
The trigger system automatically switches from "Idle" to
"Waiting for Trigger."
:TRIG
How instrument is triggered
The instrument is automatically triggered within itself.
250).