Error Messages
Error Messages
T
-223
Too many digits
The received numeric value is too long (exceeds 255 digits).
-223
Too much data
The block-, expression-, or character-string-type program data that are
received conform with the standard but exceed the amount that can be
processed under the condition of the memory or conditions specific to
memory-related devices. In this instrument, this error occurs when the number
of characters exceeds 254 in a character-string parameter.
-211
Trigger ignored
The instrument receives and detects a trigger command (":TRIG") or an
external trigger signal, but it is ignored due to the timing condition (the
instrument is not in the wait-for- trigger state, for example). Change the setup
so that a trigger command or an external trigger signal can be sent after the
instrument has entered the wait-for- trigger state.
U
-113
Undefined header
A command not defined in this instrument, though not illegal in the syntactic
structure, has been received. For example, when the message
":OUTP:DC:ISOL:FREQ:AUTO ON" is sent instead of the correct program
message ":OUTP:DC:ISOL:LEV:AUTO ON" the message sent is received as an
undefined command by this instrument. See the command reference and use
the correct command.
V
14
V bias disabled
The DC voltage bias is set when the 42841A is connected. Set the DC current
bias or disconnect the 42841A.
45
Valid in single mode only
SINGLE channel correction data is measured when the MULTI channel
correction mode is set.
Keysight E4980A/AL User's Guide