Configuring Measurement Conditions (Display and Function Related Settings)
LIST SWEEP SETUP Page

Sweep Points and Limit Modes

Functional Description
The list sweep measurement feature supports up to 201 sweep points as well
as measurement limit values.
configure the sweep points, limit parameters, and upper and lower limit values.
Figure 4-22
Sweep point fields
1
2
3
4
If you configure only the lower limit value for a sweep point, DUTs that are
below the lower limit value for the selected primary or secondary
parameter are evaluated as LOW. If you configure only the upper limit
value for a sweep point, DUTs that are above the upper limit value for the
selected primary or secondary parameter are evaluated as HIGH.
If you specify the limit parameter for a sweep point but omit both lower
and upper limit values, DUTs always pass the sweep point evaluated as IN.
146
sweep points
limit parameters
lower limit values
upper limit values
Figure 4-22
shows the fields you can use to
Keysight E4980A/AL Precision LCR Meter