Keysight E4980A User Manual page 489

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Scanner Interface
Signal Input/Output Connector
Table F-7
Scanner interface I/O connector pin assignments
Pin No.
Signal Name
8
/CH1
9
/CH3
10
/CH5
11
/CH7
12
EXT_TRIG
13
/EOM
14
COMMON
Assertion timing for these signals is different from normal measurements
and list sweep measurements.
The scanner system will operate more efficiently when the following
input/output control signals are used:
Step 1. Set the scanner channel to CHANNEL 0, the first scanner channel.
Step 2. Set the channel selection signals (/CH0 to /CH7) and the channel valid signal
(/CH_VALID) to compensate CHANNEL 0.
Step 3. Trigger the E4980A/AL with a trigger pulse input through the scanner
input/output connector on the rear panel The E4980A/AL will acknowledge
the channel number as CHANNEL0 and then measure the device connected
to scanner channel 0 using the correction data for it.
Step 4. Disassert /CH_VALID after disasserting /INDEX.
Step 5. Set the scanner channel, channel selection signals, and the channel valid
signal to the next channel to be measured when /INDEX is asserted.
Step 6. After /EOM is asserted, the E4980A/AL is ready to make the next
measurement.
Step 7. Repeat
Keysight E4980A/AL User's Guide
Input/Output
Input
Input
Output
-
Step 4
through
Step
6.
Description
Channel Selection Signal
External Trigger Signal
End-of-Measurement Signal
Common for EXT. DCV
-489