Configuring Measurement Conditions (Display and Function Related Settings)

LIMIT TABLE SETUP Page

LIMIT TABLE SETUP Page
Pressing the [Meas Setup] key followed by the LIMIT TABLE softkey opens the
LIMIT TABLE SETUP page. The LIMIT TABLE SETUP page allows you to
configure the E4980A/AL's built-in comparator.
The built-in comparator can sort DUTs into a maximum 10 levels (BIN1 through
BIN9 and OUT OF BINS) using up to nine sets of primary parameter limits
along with one set of secondary parameter limits. In addition, DUTs that fall
within the primary parameter limits but are out of the secondary parameter
limits can be sorted into an auxiliary bin.
The comparator counts DUTs up to 999999. When this number is reached, the
screen shows "-----" and the counting stops.
To take full advantage of the comparator, we recommend that you equip the
E4980A/AL with the handler interface option for use in conjunction with the
comparator. The limits for bin sorting can only be configured on the LIMIT
TABLE SETUP.
In this page, you can configure each of the following controls with the cursor
placed in the corresponding field (denoted in parentheses).
Figure 4-14
corresponding to them.
130
Measurement function (FUNC field)
Comparator limit mode (MODE field)
Tolerance mode nominal value (NOM field)
Auxiliary bin ON/OFF (AUX field)
Comparator ON/OFF (COMP field)
Beep feature (BEEP field)
Low limit for each bin (LOW field)
High limit for each bin (HIGH field)
shows the fields available on this page along with the softkeys
Keysight E4980A/AL Precision LCR Meter