Overview of Remote Control
Status Byte
Table 8-4
Standard Event Status Register Assignments
Bit
Bit
Name
No.
Weight
0
1
Operation Complete (OPC)
Bit
1
2
Request Control (RQC) Bit
2
4
Query Error (QYE) Bit
3
8
Device specific error (DDE)
Bit
4
16
Execution Error (EXE) Bit
5
32
Command Error (CME) Bit
6
64
User Request (URQ) Bit
7
128
Power On (PON) Bit
274
Description
This bit is set to "1" when the
selected pending operations before sending the *OPC
command.
always 0 (zero)
This bit is set to "1" when reading data from the output
buffer and no data was present, or when the data was lost.
This bit is set to "1" when a device-dependent error
(except for the command error, query error, and execution
error) has occurred.
This bit is set to "1" when a parameter following a header
of the GPIB command was evaluated by the
being outside its legal input range or is otherwise
inconsistent with the
E4980A/AL
This bit is set to "1" if the following command errors occur.
•
IEEE 488.2 syntax error
•
The device received a TRIGGER (GET) bus
command inside a program message.
always 0 (zero)
This bit is set to "1" when the
been turned off and then on since the last time this
register was read.
Keysight E4980A/AL Precision LCR Meter
has completed all
E4980A/AL
as
E4980A/AL
's capabilities.
's power supply has
E4980A/AL