Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
During the measurement, the ABORT softkey is shown. Use this key when
you want to abort open correction
Step 10. Repeat
measurement points.
Step 11. Move the cursor into the OPEN field.
Step 12. Press the ON key to enable the open correction feature.
To specify frequency points and carry out short correction:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the SPOT No. field.
Step 4. Enter the measurement point number using the softkeys or entry keys. If you
have used the entry keys to enter the value, the softkey labels change to unit
labels (x1).
You can specify up to 201 measurement points.
Softkey
INCR ++
INCR +
DECR -
DECR --
Step 5. Using the cursor keys, select the FREQ field.
Step 6. Enter the frequency using the entry keys. When you enter the frequency, the
softkey labels change to unit labels (Hz, kHz, MHz).
Step 7. Press the ON softkey.
Step 8. Connect the UNKNOWN terminal and the test fixture and short-circuit the
HIGH and LOW test terminals.
Step 9. Press the MEAS SHORT softkey.
During the measurement, a "SHORT measurement in progress" message is
shown on the display.
When the measurement has finished, the "SHORT measurement in
progress" message disappears.
During the measurement, the ABORT softkey is shown. Use this key when
you want to abort short correction
Keysight E4980A/AL Precision LCR Meter
Step 3
to
Step 9
until you have specified all of your desired
Description
Increments the measurement point number in steps of 10.
Increments the measurement point number in steps of 1.
Decrements the measurement point number in steps of 1.
Decrements the measurement point number in steps of 10.
121