Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
Turning On/Off the Comparator
Functional Description
The E4980A/AL's built-in comparator can sort DUTs into a maximum 10 levels
(bin 1 through bin 9 and OUT OF BINS) using up to nine sets of primary
parameter limits along with one set of secondary parameter limits.
In addition, DUTs that fall within the primary parameter limits but are out of the
secondary parameter limits can be sorted into an auxiliary bin.
To turn on or off the comparator:
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the COMP field.
Step 4. Turn on or off the comparator by pressing the appropriate softkey:
Softkey
ON
OFF
136
Description
Turns ON the comparator.
Turns OFF the comparator.
Keysight E4980A/AL Precision LCR Meter