List:bias:current - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
Description
Sets the limit values of the limit function for the list sweep measurement. When
A or B is the parameter, both lower and the upper limit should be input. If
either of them is not input, error No. 109 is returned. In the case of
OFF,<low>,<high> are optional.
Parameter
A
B
OFF (Preset value)
Preset value
Unit
Equivalent key
[Meas Setup] - LIST SETUP - LMT - A
[Meas Setup] - LIST SETUP - LMT - B
[Meas Setup] - LIST SETUP - LMT - 
[Meas Setup] - LIST SETUP - LOW
[Meas Setup] - LIST SETUP - HIGH

:LIST:BIAS:CURRent

Syntax
:LIST:BIAS:CURRent <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:CURRent?
Description
Clears the previous list sweep table, sets DC current sweep as the list sweep
parameter, and sets DC current points for the DC current list sweep. If the list
sweep is not set for DC current, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Range of n
Unit
Keysight E4980A/AL Precision LCR Meter
Description
Sets limit test target to A (Primary parameter)
Sets limit test target to B (Secondary parameter)
Sets limit test target to OFF
<low>,<high>
Query returns -9.9E37, +9.9E37 for <low>, <high>,
respectively in the case of OFF.
Depends on
:FUNCtion:IMPedance[:TYPE]
<point n>
1 to 201
A
353

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4980aalE4980al

Table of Contents