Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
Figure 4-17
Example of sorting in tolerance mode
When the limit table contains any value and you attempt to change the
comparator limit mode, you are prompted with a "Clear the table first"
warning message. If this is the case, clear the table and then change the
limit mode. For information on how to clear the table, see
entire table:" on page
To set up the comparator limit mode:
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the MODE field.
Step 4. Select your desired limit mode by pressing the appropriate softkey:
Softkey
ABS
%
SEQ
134
142.
Description
Switches the comparator into tolerance mode based on the absolute
parameter values.
Switches the comparator into tolerance mode based on the
deviation percentages.
Switches the comparator into sequential mode.
"To clear the
Keysight E4980A/AL Precision LCR Meter