Display Format
MEAS DISPLAY Page
Conditions in which overload occurs for Vdc-Idc measurement (DC
source-level monitor)
Conditions differ depending on the Idc impedance range setting (DCI RNG
field).
Impedance
range
AUTO
HOLD
Overload does not occur for Vdc measurement.
Table 3-25
Overload range when the impedance range is in the HOLD state for Idc
measurement
Impedance range (A)
(DCI RNG field)
20
200
2 m
20 m
100 m
List of measurement results when overload occurs.
This section describes measurement results and each monitor value when
overload occurs.
The bold font in the list indicates the measurement parameter for which
overload occurs.
"*" indicates that measurement results, judgment, and BIN count are normal.
Keysight E4980A/AL Precision LCR Meter
Cond itions
•
When the measurement value exceeds 125 mA
•
When the E4980A/AL's internal detector detects overload
Conditions differ depending on the impedance range. Refer to
3-25
Measurable range
0 to 22
0 to 220
0 to 2.2 m
0 to 22 m
0 to 110 m
Table
Range inn which overload
occurs
more than 22
more than 220
more than 2.2 m
more than 22 m
more than 110 m
73