Save/Recall

Saving Measurement Results into USB Memory

Saving Measurement Results into USB Memory
You can save measurement results obtained by the E4980A/AL into USB
memory as .CSV files. You can later load your saved files into an application
program running on a PC.
You can save up to 201 sets of measurement results at a time, including
measurement results from a list sweep.
Before saving measurement results into USB memory, you need to complete
some tasks preparatory to initiating measurement so that the data buffer
memory contains the necessary data. For more information, see
measurement results into USB memory:" on page
You cannot obtain measurement results that come after your specified
value from the data buffer memory. However, you can set up the data
buffer memory using the
You cannot load measurement results from USB memory into the
E4980A/AL.

Measurement Result Format

Measurement results are output in the format shown in
The Data A and Data B fields of a measurement result file differ depending on
whether the "effective digits" setting is on or off (controlled by the
:FORM:ASC:LONG command).
Table 6-3
Measurement Result Format
Measure
ment
Screen
one point
one point
list
Data A
Outputs the measurement data for the primary parameter.
The Data A field uses the following two different fixed-length ASCII formats:
When the "effective digits"
setting is off:
(:FORM:ASC:LONG OFF)
178
":MEMory:DIM" on page 359
Comparator
Resul t Format
OFF
<Data A>,<Data B>,<Status>
ON
<Data A>,<Data B>,<Status>,<BIN No.>
ON/OFF
<Data A>,<Data B>,<Status>,<IN/OUT>
SN.NNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
181.
command.
Table
6-3.
Keysight E4980A/AL Precision LCR Meter
"To save