SCPI Command Reference
E4980A/AL commands
Parameter
ON or 1 (Preset value)
OFF or 0
Equivalent key
No equivalent key is available on the front panel.
:FUNCtion:SMONitor:IDC[:STATe]
Syntax
:FUNCtion:SMONitor:IDC[:STATe] {ON|OFF|1|0}
:FUNCtion:SMONitor:IDC[:STATe]?
Description
Enables the DC current-level monitor function.
This is available when option 001 is installed.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - MEAS SETUP - IDC MON - ON
[Meas Setup] - MEAS SETUP - IDC MON - OFF
:FUNCtion:SMONitor:VAC[:STATe]
Syntax
:FUNCtion:SMONitor:VAC[:STATe] {ON|OFF|1|0}
:FUNCtion:SMONitor:VAC[:STATe]?
Description
Does nothing. The E4980A/AL always has its AC voltage-level monitor turned
ON. This function is provided because of its command compatibility with
4284A.
Parameter
ON or 1 (Preset value)
350
Description
Does nothing
Does nothing
Description
DC current monitor function ON
DC current monitor function OFF
Description
Does nothing
Keysight E4980A/AL Precision LCR Meter