SCPI Command Reference
E4980A/AL commands
Parameter
Unit
Range of n
Unit
Example of Use
viVPrintf(AgtE4980A/AL, "COMP:MODE SEQ" + vbLf, 0)
viVPrintf(AgtE4980A/AL, "FUNC:IMP RX" + vbLf, 0)
viVPrintf(AgtE4980A/AL, "COMP:SEQ:BIN -9.9e37,
100,110,120,130,140, 9.9e37,150" + vbLf, 0)
Equivalent key
[Meas Setup] - LIMIT TABLE - Bin 1 LOW
[Meas Setup] - LIMIT TABLE - Bin 1 - 9 HIGH
:COMParator:SLIMit
Syntax
:COMParator:SLIMit <Low>,<High>
:COMParator:SLIMit?
Description
Sets the lower/upper limit values for the comparator function's secondary
parameter.
Parameter
Preset value
Unit
Equivalent key
[Meas Setup] - LIMIT TABLE - 2nd LOW
[Meas Setup] - LIMIT TABLE - 2nd HIGH
:COMParator[:STATe]
Syntax
:COMParator[:STATe] {ON|OFF|1|0}
:COMParator[:STATe]?
Description
Enables the comparator function.
328
<Bin 1 Low>
Depends on
:FUNCtion:IMPedance[:TYPE]
<Bin n High>
1 to 9
Depends on
:FUNCtion:IMPedance[:TYPE]
<Low>,<High>
-9.9E+37, 9.9E+37
Depends on
:FUNCtion:IMPedance[:TYPE]
Keysight E4980A/AL Precision LCR Meter