Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
1. This area is used to:
2. This area is used to:
The following tables show how correction data elements are related to each
other when the correction features are enabled in the OPEN, SHORT, and LOAD
fields. Correction data elements are selected based on the number of
measurement points.
•
Correction
OPEN: ON
SHORT: ON
LOAD: ON
•
Correction
OPEN: ON
SHORT: ON
LOAD: ON
User-specifie
d
Interpolated
124
•
Carry out correction calculation using either the open/short correction
data generated by the interpolation method or the open/short/load
correction data at specified frequency points.
Which data is used depends on the test frequencies.
•
Obtain the open/short correction data.
•
Obtain the open/short/load correction data at each frequency point
specified in the FREQ field.
•
Enable or disable the open/short/load correction data at each frequency
point specified in the FREQ field.
When test frequency point is equal to the frequency point specified in FREQ
field:
Frequency point specified in FREQ field
ON
User-specified
User-specified
User-specified
When test frequency point is NOT equal to the frequency point specified in
FREQ field:
Frequency point specified in FREQ field
ON
Interpolated
Interpolated
X
Correction data at your specified frequency point are used.
Correction data generated by interpolation are used.
OFF
Interpolated
Interpolated
X
OFF
Interpolated
Interpolated
X
Keysight E4980A/AL Precision LCR Meter