Measurement Procedure and Examples
Measurements Using DC source
6. Move to the SHORT field using the cursor keys and press the MEAS SHORT
7. Press the ON softkey to enable a SHORT correction.
Step 6. Connect the transistor base and DC source. The corrector and emitter are
connected to the DUT.
Figure 7-24
Connecting DUT
Step 7. Press the [Meas Setup] key to set the following measurement conditions in
each field of the LIST SWEEP SETUP page
1. Set the list sweep parameter to DC SRC[V].
2. Enter the level of the sweep point.
Step 8. Press the [DC BIAS] key. DCSRC is displayed in the status display area and
the LED indicator of the DC source is ON (orange).
218
softkey.
Wait until the message "SHORT measurement in progress" disappears.