SCPI Command Reference
E4980A/AL commands
[Meas Setup] - MEAS SETUP - DCR RNG - HOLD

:FUNCtion:DCResistance:RANGe[:VALue]

Syntax
:FUNCtion:DCResistance:RANGe[:VALue] <numeric>
:FUNCtion:DCResistance:RANGe[:VALue]?
Description
Selects the DCR measurement range. This command turns the auto range
function OFF.
Parameter
Range
Preset value
Unit
Equivalent key
[Meas Setup] - MEAS SETUP - DCR RNG

:FUNCtion:DEV[1-2]:MODE

Syntax
:FUNCtion:DEV[1-2]:MODE {ABSolute|PERCent|OFF}
:FUNCtion:DEV[1-2]:MODE?
Description
Selects the deviation measurement mode.
Parameter
ABSolute
PERCent
OFF(Preset value)
Equivalent key
[Meas Setup] - MEAS SETUP - DEV A - ABS
[Meas Setup] - MEAS SETUP - DEV A - %
[Meas Setup] - MEAS SETUP - DEV A - OFF
[Meas Setup] - MEAS SETUP - DEV B - ABS
[Meas Setup] - MEAS SETUP - DEV B - %
[Meas Setup] - MEAS SETUP - DEV B - OFF
346
<Numeric>
10|100|1k|10k|100k
100
ohms
Description
Sets the deviation measurement mode to "ABS"
Sets the deviation measurement mode to "%"
Disables the deviation measurement mode
Keysight E4980A/AL Precision LCR Meter