SCPI Command Reference
E4980A/AL commands
Parameter
ON or 1 (Preset value)
OFF or 0
Equivalent key
[Display Format] - DISPLAY BLANK

:DISPlay:LINE

Syntax
:DISPlay:LINE <String>
:DISPlay:LINE?
Description
Enters arbitrary comments containing up to 30 ASCII characters in the
comment field. The string "USER COMMENT" is displayed if this is empty.
Parameter
Preset value
Equivalent key
[Meas Setup] - MEAS SETUP - USER COMMENT - ADD CHAR
[Meas Setup] - MEAS SETUP - USER COMMENT - ENTER
[Meas Setup] - MEAS SETUP - USER COMMENT - NEXT
[Meas Setup] - MEAS SETUP - USER COMMENT - PREV

:DISPlay:PAGE

Syntax
:DISPlay:PAGE
{MEASurement|BNUMber|BCOunt|LIST|MSETup|CSETup|LTABle|LSETup|CATAlo
g|SYSTem|SELF|MLARge|SCONfig|SERVice}
:DISPlay:PAGE?
Description
Selects the page to be displayed.
Parameter
MEASurement (Preset value)
BNUMber
340
Description
Enables display updates
Disables display updates. The display will be blank.
<String>
"" (NULL)
Description
Sets displayed page to <MEAS DISPLAY>
Sets displayed page to <BIN No. DISPLAY>