Handler Interface
Setting Up the Handler Interface Board
Figure E-9
Flow chart for determining the jumper and bit switch settings
Step 6. Set up the jumpers as instructed in
Table E-11
Jumper and bit switch settings for test result output signals
Setting #
JP1
Setting 1
1
Setting 2
2
Setting 3
3
Setting 4
-
-476
Bit 5 of
Circuit common
S1
On
Instrument's circuit
common
On
Instrument's circuit
common
Off
COM1
Off
COM1
.
Table E-11
Pull-up resistor
mounting
Required
Required
Required
+5 V through +24 V,
Not required
-
Keysight E4980A/AL User's Guide
Pull-up vol tage
+5 V, internal
+12 V, internal
external
(EXT.DCV1)