Scanner Interface
How to Use the Scanner Interface Board
standard capacitor that has reference parallel capacitance and D values, the
reference values should be stored with the Cp-D function. The reference values
are also stored on the CORRECTION page.
The function of the standard on the CORRECTION page can be defined
independently of the measurement function set on the MEAS SETUP page.
For example, if the reference values are defined as the Cp-D function
(entered with capacitance value and D value), the Z- parameter can be
measured while performing the correction.
The setting procedure is given below.
Step 1. Press the [Meas Setup] - CORRECTION to display the CORRECTION page.
Step 2. Use the cursor arrow keys on the front panel to move the cursor to the FUNC
field.
Step 3. Use the softkeys to select the function to be used for entering the reference
values.
Step 4. Use the cursor arrow keys on the front panel to move the cursor to the SPOT
No. field.
Step 5. Use the numeric entry keys or softkeys on the front panel to enter a
measurement point number for entering the reference value.
Step 6. Use the cursor arrow keys on the front panel to move the cursor to the FREQ
field.
Step 7. When the FREQ field is OFF, press the ON softkey to use the correction data
of the measurement point number.
Step 8. Use the numeric entry keys and softkeys on the front panel to enter a test
frequency at a specific measurement point.
Step 9. Use the cursor arrow keys on the front panel to move the cursor to the REF A
field.
Step 10. Use the numeric entry keys and softkeys on the front panel to enter the
reference value of the primary parameter at the specified measurement point.
Step 11. Use the cursor arrow keys on the front panel to move the cursor to the REF B
field.
Step 12. Use the numeric entry keys and softkeys on the front panel to enter the
reference value of the secondary parameter at the specified measurement
point.
Keysight E4980A/AL User's Guide