SCPI Command Reference
E4980A/AL commands
Description
Enables automatic IP address setup. The E4980A/AL can obtain an available IP
address at 169.254.xxx.xxx.
This command works with the firmware revision 1.0x. Do not use this
command with the firmware revision 2.00 or later.
Parameter
ON or 1 (Preset value)
OFF or 0
Equivalent key
[System] - SYSTEM CONFIG - AUTO-IP ENABLED - ON
[System] - SYSTEM CONFIG - AUTO-IP ENABLED - OFF

:SYSTem:COMMunicate:LAN[:SELF]:CONFigure

Syntax
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure {AUTO|MANual}
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure?
Description
Selects automatic obtaining or manual setting for the IP address obtaining
method.
This command is available for the firmware revision 2.00 or later.
Parameter
AUTO (Preset value)
MANual
Equivalent key
[System] - SYSTEM CONFIG - IP CONFIG - AUTO
[System] - SYSTEM CONFIG - IP CONFIG - MAN

:SYSTem:COMMunicate:LAN[:SELF]:CONTrol

Syntax
:SYSTem:COMMunicate:LAN[:SELF]:CONTrol
Description
Returns the SOCKET control port number. If the parser is a SOCKET, it will
return a number from 5000 to 5100. Otherwise, 0. (Query Only)
Keysight E4980A/AL Precision LCR Meter
Description
Enables automatic IP address setup
Disables automatic IP address setup
Description
Sets the IP address obtaining method to AUTO
Sets the IP address obtaining method to MAN