Display Format
MEAS DISPLAY Page
Effective measurement range for each impedance range for impedance
measurement
Figure 3-3
range for each impedance range while in the impedance mode (| Z |, R, X). For
example, 50-k DUT impedance can be measured using the ranges of 100 k
to 30 k, but the E4980A/AL's accuracy specification is only met when using
the 30-k range. If this DUT is measured with the 100-k range, "OVERLOAD"
will be displayed.
Figure 3-3
Effective measurement range for each impedance range (Impedance
measurement)
Table 3-6
Effective measurement range for the impedance range when it is in the HOLD
state for impedance measurement
Impedance range ()
100 m
1
10
100
300
1 k
3 k
52
shows the range that can be displayed and the effective measuring
Effective measurement range
0 < Impedance 108 m
1
108 m
1.08 < Impedance 10.08
10.08 < Impedance < 276
276 Impedance < 0.92 k
0.92 k Impedance < 2.76 k
2.76 k Impedance < 9.2 k
2
< Impedance 1.08
Keysight E4980A/AL Precision LCR Meter