Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
Figure 4-8
CORRECTION page

To set the correction function to on or off

Open/short correction at all frequency points and open/short/load correction
at user-specified spot frequency points can be enabled or disabled by entering
ON or OFF (with softkeys) in the OPEN, SHORT and LOAD fields on the
CORRECTION display page. The on/off settings in the OPEN, SHORT and
LOAD fields determine the on/off state of both open/short correction at all
frequency points and open/short/load correction at user-specified spot
frequency points. Accordingly, open/short/load correction at user-specified
spot frequency points can be performed only when open/short correction at all
frequency points is activated.
When the settings of the OPEN, SHORT and LOAD fields are ON, both
open/short correction at all frequency points and open/short/load correction
at user-specified spot frequency points are enabled. Open/short/load
correction at user-specified spot frequency points are not carried out by setting
FREQ field (below SPOT NO. field) to default OFF state. By entering a desired
frequency in the FREQ field allocated to SPOT NO. from 1 to 201,
open/short/load correction at the user-specified spot frequency is carried out.
Open/short /load correction at user-specified spot frequency points takes
priority over open/short correction at all frequency points when test frequency
is identical with one of the user-specified spot frequency points.
Keysight E4980A/AL Precision LCR Meter
113