Comparator:swap; Comparator:tolerance:bin[1-9] - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - LIMIT TABLE - COMP - ON
[Meas Setup] - LIMIT TABLE - COMP - OFF

:COMParator:SWAP

Syntax
:COMParator:SWAP {ON|OFF|1|0}
:COMParator:SWAP?
Description
Enables the swap parameter function.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - LIMIT TABLE - FUNC - SWAP PARAM

:COMParator:TOLerance:BIN[1-9]

Syntax
:COMParator:TOLerance:BIN[1-9] <low>,<high>
:COMParator:TOLerance:BIN[1-9]?
Description
Sets the lower/upper limit values of each BIN for the comparator function's
tolerance mode. These limits can be set only when the limit mode is set to the
tolerance mode. When the lower limit or upper limit is not set, it returns
-9.9E37 and 9.9E37, respectively.
Parameter
Unit
Equivalent key
[Meas Setup] - LIMIT TABLE - LOW
[Meas Setup] - LIMIT TABLE - HIGH
Keysight E4980A/AL Precision LCR Meter
Description
Enables the comparator function
Disables the comparator function
Description
9 BIN settings are used for the secondary parameter
9 BIN settings are used for the primary parameter
<low>,<high>
Depends on
:FUNCtion:IMPedance[:TYPE]