Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page
Table 4-5
Resolution steps when setting the trigger delay time with INCR + / DECR -
Trigger delay time
0 s through 100 ms
100 ms through 1 s
1 s through 10 s
10 s through 100 s
100 s through 999 s
Figure 4-5
Trigger delay time and step delay time
In DCR measurement, there are two step delay times because two cycles
of measurement are performed by applying +0.1 V and -0.1 V. Accordingly,
when measuring the Rdc parameter such as Ls-Rdc, since the stimulus
setting and measurement procedure is executed three times in one
measurement, the step delay time takes three times as long as the setting
value.
102
Resolution
100 s
1 ms
10 ms
100 ms
1 s