Display Format
MEAS DISPLAY Page
Table 3-18
DC bias and resolution (Option 001)
DC bias signal current level
(0A to 50 mA)
(50 mA to 100 mA)
When a DUT is connected to the measurement contacts, the setting current
value is different from the actual current passing through the DUT. For normal
measurement taken by applying the signal current through a device, refer to
"DC Bias Current Isolation" on page
Figure 3-7
DC bias signal current
Setting limits for DC bias and test signal level
When both the DC bias and the test signal level are set under the following
conditions, the amount of DC bias plus the test signal level is limited to the
values listed in
•
•
Table 3-19
DC bias and test signal level-setting limits
Setting value
DC bias
Vdc (V)
Vdc (V)
Keysight E4980A/AL Precision LCR Meter
1. Effective resolution is 3.3 A.
Table 3-19
Option 001 is installed.
The DC Bias key on the front panel is set to ON.
Test signal
level
Vosc (Vrms)
Iosc (Arms)
Resolution
1
1 A
10 A
97.
Limit
Vosc 2 1.15 + Vdc 1.002 < 42 V
Iosc 2 115 + Vdc 1.002 < 42 V
65