Display Format
MEAS DISPLAY Page
Test signal level and resolution (Option 001)
When option 001 (power/DC bias enhance) is installed, the oscillator voltage
level can be set with a resolution listed in
level can be set with a resolution listed in
Table 3-12
Oscillator voltage level and resolution (Option 001)
Oscillator vol tage level
0 Vrms to 200 mVrms
200 mVrms to 500 mVrms
500 mVrms to 1 Vrms
1 Vrms to 2 Vrms
2 Vrms to 5 Vrms
5 Vrms to 10 Vrms
10 Vrms to 20 Vrms
Table 3-13
Oscillator current level and resolution (Option 001)
Oscillator current level
0 Arms to 2 mArms
2 mArms to 5 mArms
5 mArms to 10 mArms
10 mArms to 20 mArms
20 mArms to 50 mArms
50 mArms to 100 mArms
Test signal level setting procedure
Step 1. Press the [Display Format] key
Step 2. Press the MEAS DISPLAY softkey.
Step 3. Use the cursor keys to select the LEVEL field.
62
1
1. When the test frequency is more than 1 MHz, the maximum oscillator voltage
level that can be set is 15 Vrms.
Table
3-12, or the oscillator current
Table
3-13.
Resolution
100 Vrms
200 Vrms
500 Vrms
1 mVrms
2 mVrms
5 mVrms
10 mVrms
Resolution
1 Arms
2 Arms
5 Arms
10 Arms
20 Arms
50 Arms
Keysight E4980A/AL Precision LCR Meter