Keysight E4980A User Manual page 120

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
Figure 4-12
Open/short/load correction
To specify frequency points and carry out open correction:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the SPOT No. field.
Step 4. Enter the measurement point number using the softkeys or entry keys. If you
have used the entry keys to enter the value, the softkey labels change to unit
labels (x1).
You can specify up to 201 measurement points.
Softkey
INCR ++
INCR +
DECR -
DECR --
Step 5. Using the cursor keys, select the FREQ field.
Step 6. Enter the frequency using the entry keys. When you have entered the
frequency, the softkey labels change to unit labels (Hz, kHz, MHz).
Step 7. Press the ON softkey.
Step 8. Connect the UNKNOWN terminal and the text fixture with no DUT connected.
Step 9. Press the MEAS OPEN softkey.
120
Description
Increments the measurement point number in steps of 10.
Increments the measurement point number in steps of 1.
Decrements the measurement point number in steps of 1.
Decrements the measurement point number in steps of 10.
During the measurement, an "OPEN measurement in progress" message is
shown on the display.
When the measurement has finished, the "OPEN measurement in progress"
message disappears.
Keysight E4980A/AL Precision LCR Meter