Handler Interface
Setting Up the Handler Interface Board
where:
— Vp [V]: pull-up voltage
— R [k]: pull-up resistance
For the typical pull-up resistor values, refer to
Step 11. Set up the bit switch for input signals as instructed in
Table E-13
Bit switch settings for input signals
Drive vol tage
+5 V, internal
+12 V, internal
+5 V through +6 V,
external
(EXT.DCV2)
+6 V through +9 V,
external
(EXT.DCV2)
+9 V through +15 V,
external
(EXT.DCV2)
+15 V through +24 V,
external
(EXT.DCV2)
Step 12. Connect the flat cable to the handler interface board. Reinstall the handler
interface board to its original position and secure it with the two screws.
When tightening each screw, take care not to apply excessive force
because doing so can break the tapped hole. (0.98 N-Em, 0.1 kgf-m Max).
The typical pull-up resistor values are:
Table E-14
The typical pull-up resistor vaules
Pull-Up Vol tage
5 V
-478
Bits 1 through 4 of
1
S1
Turn ON bit 1.
Turn ON bit 3.
Turn ON bit 1.
Turn ON bit 2.
Turn ON bit 3.
Turn ON bit 4.
1. Only one of bits 1 through 4 of the bit switch can be ON at a time.
Pull-Up Resistor
1.7 k ( = 5 V / 3 mA)
Table E-14 on page
Table
E-13.
Circuit common
Instrument's circuit
common
Instrument's circuit
common
COM2
COM2
COM2
COM2
Keysight Part Number
1810-0276 (1.5 k)
Keysight E4980A/AL User's Guide
478.