Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
Step 10. Repeat
measurement points.
Step 11. Move the cursor into the SHORT field.
Step 12. Press the ON key to enable the short correction feature.
To specify frequency points and carry out load correction:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the SPOT No. field.
Step 4. Enter the measurement point number using the softkeys or entry keys. If you
use the entry keys to enter the value, the softkey labels change to unit labels
(x1).
You can specify up to 201 measurement points.
Softkey
INCR ++
INCR +
DECR -
DECR --
Step 5. Using the cursor keys, select the FREQ field.
Step 6. Enter the frequency using the entry keys. When you enter the frequency, the
softkey labels change to unit labels (Hz, kHz, MHz).
Step 7. Press the ON softkey.
Step 8. Move the cursor into the REF A field.
Step 9. Enter the known value of the standard's primary parameter using the softkeys
or entry keys.
Step 10. Move the cursor into the REF B field.
Step 11. Enter the known value of the standard's secondary parameter using the
softkeys or entry keys.
Step 12. Move the cursor into the FREQ field.
Step 13. Connect the standard to the UNKNOWN terminal.
Step 14. Press the MEAS LOAD softkey.
122
Step 3
to
Step 9
until you have specified all of your desired
Description
Increments the measurement point number in steps of 10.
Increments the measurement point number in steps of 1.
Decrements the measurement point number in steps of 1.
Decrements the measurement point number in steps of 10.
During the measurement, a "LOAD measurement in progress" message is
shown on the display.
Keysight E4980A/AL Precision LCR Meter