Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page
Averaging Factor
Functional Description
The averaging feature of the E4980A/AL allows you to obtain moving average
values of successive measurement results.
You can specify the averaging factor within the range of 1 to 256 in steps of 1.
The averaging feature does not affect the measurement results of the DC
bias voltage monitor and DC bias current monitor even when a particular
averaging factor is specified.
To set up the averaging factor:
Step 1. Press [Meas Setup].
Step 2. Using the cursor keys, select the AVG field.
Step 3. Enter the averaging factor using the softkeys or entry keys. If you have used
the entry keys to enter the value, the softkey labels change to unit labels (x1).
Softkey
INCR ++
INCR +
DECR -
DECR --
98
Description
Increments the averaging factor in steps of 1, 2, 4, 8, 16, 32, 64,
128, and 256.
Increments the averaging factor in steps of 1.
Decrements the averaging factor in steps of 1.
Decrements the averaging factor in steps of 1, 2, 4, 8, 16, 32, 64,
128, and 256.
Keysight E4980A/AL Precision LCR Meter