Bias:current[:Level] - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
Parameter
SHORt
MEDium (Preset value)
LONG
Range
Preset value
Resolution
Equivalent key
[Meas Setup] - MEAS SETUP - MEAS TIME - SHORT
[Meas Setup] - MEAS SETUP - MEAS TIME - MED
[Meas Setup] - MEAS SETUP - MEAS TIME - LONG
[Meas Setup] - MEAS SETUP - AVE

:BIAS:CURRent[:LEVel]

Syntax
:BIAS:CURRent[:LEVel] <numeric>
:BIAS:CURRent[:LEVel]?
Description
Sets the DC bias current. Setting does not implicitly turn the DC bias ON. This
command is effective when option 001is installed. Without Option 001, this
command causes an error. If the bias is set by :BIAS:VOLTage[:LEVel], the
query command returns error No. 230. The parameter resolution depends on
the current setting value.
Parameter
Range
Unit
Equivalent key
[Display Format] - MEAS DISPLAY - BIAS
[Meas Setup] > - MEAS SETUP - BIAS
322
1. Depending on signal level.
Description
Sets measurement time to Short
Sets measurement time to Medium
Sets measurement time to Long
<Numeric>
1 to 256
1
1
<Numeric>
1
-100 m to 100 m
(with option 001)
A
Keysight E4980A/AL Precision LCR Meter

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4980aalE4980al

Table of Contents